Delivery Limit values should be displayed on the Settings page

  • Having a Delivery Limits value of "Enabled" is (basically) useless
  • Loop dev and presumably Tidepool Loop already do this
  • Showing actual values consistent with all other configuration settings

Delivery Limits should be the closed loop delivery limits, not necessarily the pump's programmable delivery limits

  • Some pump types have no hardware settable delivery limits (e.g, Omnipod), just design enforced delivery limits
  • Pumps with settable delivery limits (e.g., Medtronic) just need compatible delivery limits. E.g., A user should be able to
    selectively decrease the Loop max basal limit w/o the pump enforced delivery limit changing. Only when the Loop delivery limit
    is increased to a value greater than the current pump maximum does this pump enforced limit need to be updated to match.
  • Allows for temporarily setting max basal rates lower than even the scheduled basal rates if needed/desired
    (Medtronic pump's max basal rate must be >= the highest scheduled basal rate).
  • Provides better consistency across different pump types

Delivery Limits should include a minimum basal rate

  • Gives advanced users/athletes more control
  • Relatively obvious behavior
  • Many users have requested to have over the years
  • Consistent behavior with Delivery Limits being a closed loop delivery limit

Additional Details & Background


When doing mountain hiking, I find that having a starting max basal rate of slightly less than my max scheduled basal rate gives my good results. Over the course of the hike, I adjust up my max basal to eventually be my normal max works best (typically in 2-4 steps over the course of many hours). I have been doing this since I started with Loop on a Medtronic pump in 2016. I was no longer able to use this technique with the PumpManager rewrite for the Omnipod. At this time, I had to modify the Medtronic PumpManager code not to always force the Medtronic delivery limit to match a lowered Loop max basal rate. With the Omnipod, has never been a problem because there is no way to program a delivery limit on a Omnipod.

When doing mountain hiking, I find that having a minimal basal rate of 0.1 U/hr throughout the hike has been helpful (it prevents me from potentially getting absolutely no insulin for an extended period). This is not possible without having a minimum basal rate. I have talked with a number of marathoners who also require a very small, but non-zero, min basal rate during races or bad things can result.

Example Screenshots


IMG_6491

IMG_6492

Note that one screen is used for setting all of the delivery limits to match previous behavior.