PyVISA

GPIB Operate Scientific Instruments for Keithley 2410

Keithley 2410 Source Meter commands

Calibration unlocked states

Mode State Equivalent remote command
Concurrent Functions = OFF :SENS:FUNC:CONC OFF
Sense Function = Source :SENS:FUNC <source_ function>
Sense Volts NPLC = 1.0 :SENS:VOLT:NPLC 1.0
Sense Volts Range = Source V :SENS:VOLT:RANG <source_V_range>
Sense Current NPLC = 1.0 :SENS:CURR:NPLC 1.0
Sense Current Range = Source I :SENS:CURR:RANG <source_ I_ range>
Filter Count = 10 :SENS:AVER:COUN 10
Filter Control = REPEAT :SENS:AVER:TCON REPeat
Filter Averaging = ON :SENS:AVER:STAT ON
Source V Mode = FIXED :SOUR:VOLT:MODE FIXED
Volts Autorange = OFF :VOLT:RANG:AUTO OFF
Source I Mode = FIXED :SOUR:CURR:MODE FIXED
Current Autorange = OFF :SOUR:CURR:RANG:AUTO OFF
Autozero = ON :SYST:AZERO ON
Trigger Arm Count = 1 :ARM:COUNT 1
Trigger Arm Source = Immediate :ARM:SOUR IMMediate
Trigger Count = 1 :TRIG:COUNT 1
Trigger Source = Immediate :TRIG:SOUR IMMediate